Repositioning and Extraction of Stylet-Driven Pacing Leads with Extendable Helix used for Left Bundle Branch Area Pacing

التفاصيل البيبلوغرافية
العنوان: Repositioning and Extraction of Stylet-Driven Pacing Leads with Extendable Helix used for Left Bundle Branch Area Pacing
المؤلفون: Bert Geerts, Jean Yves Wielandts, Rene Tavernier, Sébastien Knecht, Antonio Sorgente, Jean-Benoît le Polain de Waroux, Gabriela Hilfiker, Kris Gillis, Lucio Capulzini, Mattias Duytschaever
بيانات النشر: Authorea, Inc., 2021.
سنة النشر: 2021
مصطلحات موضوعية: medicine.medical_specialty, business.industry, Computer science, Acute complication, 030204 cardiovascular system & hematology, Stylet, 03 medical and health sciences, 0302 clinical medicine, Physiology (medical), Internal medicine, Helix, Left bundle branch, medicine, Cardiology, 030212 general & internal medicine, Cardiology and Cardiovascular Medicine, business, Lead (electronics)
الوصف: Conventional stylet-driven leads with extendable helix can be implanted successfully for Left Bundle Branch Area Pacing (LBBAP) with a low acute complication rate. However, removal and repositioning techniques of these leads haven’t been yet described. We report 2 cases in which lead repositioning after a first unsuccessful attempt to LBBAP was associated with fracture of the helix rotating mechanism and failure to fully extract the pacing lead.
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::5262c4355d79a2bde6752413fe806dc8
https://doi.org/10.22541/au.161150907.78297540/v1
حقوق: OPEN
رقم الأكسشن: edsair.doi.dedup.....5262c4355d79a2bde6752413fe806dc8
قاعدة البيانات: OpenAIRE