Dual energy CT for a small animal radiation research platform using an empirical dual energy calibration

التفاصيل البيبلوغرافية
العنوان: Dual energy CT for a small animal radiation research platform using an empirical dual energy calibration
المؤلفون: Manuela Anna Duda, Andreas Grad, Severin Kampfer, Sophie Dobiasch, Stephanie Elisabeth Combs, Jan Jakob Wilkens
المصدر: Phys. Med. Biol. 67:135009 (2022)
بيانات النشر: IOP Publishing, 2021.
سنة النشر: 2021
مصطلحات موضوعية: Mice, Radiological and Ultrasound Technology, Phantoms, Imaging, Paper, small animal, computed tomography, SARRP, CT, dual energy, spectral, imaging, Calibration, Animals, Contrast Media, Small Animal, Computed Tomography, Sarrp, Ct, Dual Energy, Spectral, Imaging, Radiology, Nuclear Medicine and imaging, Tomography, X-Ray Computed, Iodine, ddc
الوصف: Objective. Dual energy computed tomography (DECT) has been shown to provide additional image information compared to conventional CT and has been used in clinical routine for several years. The objective of this work is to present a DECT implementation for a Small Animal Radiation Research Platform (SARRP) and to verify it with a quantitative analysis of a material phantom and a qualitative analysis with an ex-vivo mouse measurement. Approach. For dual energy imaging, two different spectra are required, but commercial small animal irradiators are usually not optimized for DECT. We present a method that enables dual energy imaging on a SARRP with sequential scanning and an Empirical Dual Energy Calibration (EDEC). EDEC does not require the exact knowledge of spectra and attenuation coefficients; instead, it is based on a calibration. Due to the SARRP geometry and reconstruction algorithm, the calibration is done using an artificial CT image based on measured values. The calibration yields coefficients to convert the measured images into material decomposed images. Main results. To analyze the method quantitatively, the electron density and the effective atomic number of a material phantom were calculated and compared with theoretical values. The electron density showed a maximum deviation from the theoretical values of less than 5% and the atomic number of slightly more than 6%. For use in mice, DECT is particularly useful in distinguishing iodine contrast agent from bone. A material decomposition of an ex-vivo mouse with iodine contrast agent was material decomposed to show that bone and iodine can be distinguished and iodine-corrected images can be calculated. Significance. DECT is capable of calculating electron density images and effective atomic number images, which are appropriate parameters for quantitative analysis. Furthermore, virtual monochromatic images can be obtained for a better differentiation of materials, especially bone and iodine contrast agent.
وصف الملف: application/pdf
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::600d89d23a8f852bd7aa1b51e9abd622
https://mediatum.ub.tum.de/1663010
حقوق: OPEN
رقم الأكسشن: edsair.doi.dedup.....600d89d23a8f852bd7aa1b51e9abd622
قاعدة البيانات: OpenAIRE