In vitro evaluation of endovascular stents to assess suitability for endovascular graft fixation

التفاصيل البيبلوغرافية
العنوان: In vitro evaluation of endovascular stents to assess suitability for endovascular graft fixation
المؤلفون: D.M. Nott, A.W. Anson, S.M. Andrews, Roger M. Greenhalgh
المصدر: European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery. 9(4)
سنة النشر: 1995
مصطلحات موضوعية: medicine.medical_specialty, Swine, medicine.medical_treatment, In Vitro Techniques, Blood vessel prosthesis, medicine.artery, medicine, Alloys, Animals, cardiovascular diseases, Graft fixation, Endovascular stent, Aorta, Ultrasonography, Interventional, Fixation (histology), Medicine(all), Aortic graft, business.industry, Stent, Palmaz stent, medicine.disease, equipment and supplies, Abdominal aortic aneurysm, Surgery, Biomechanical Phenomena, Blood Vessel Prosthesis, surgical procedures, operative, Evaluation Studies as Topic, Shape Memory Alloy, Stents, Radiology, Cardiology and Cardiovascular Medicine, business, Blood Flow Velocity
الوصف: Objective: To compare and ascertain the suitability for endovascular aortic graft fixation of the Palmaz stent, the Strecker stent, the modified Gianturco-Z stent, the Wallstent and a new stent made of the shape memory alloy nitinol. Design, setting and materials: In vitro studies using a simple tensiometer to assess the forces required to distract each stent from porcine aorta; and using a high-pressure, pulsatile-flow pump to measure flow rates required to dislodge stents holding grafts within porcine aortas, with assessment using endovascular ultrasound. Results: The Palmaz, the modified Gianturco-Z and the nitinol stents resisted significantly greater distraction forces than the Wallstent, but endovascular ultrasound examination revealed that the modified Gianturco-Z stent was seen to move away from the aortic wall at high flow rates precluding adequate fixation. The Strecker stent could not be assessed using these techniques. Conclusions: A nitinol stent may be suitable for aortic graft fixation, has characteristics similar to the Palmaz stent and has shape memory effect which may overcome the difficulties of introduction of wider diameter stents through narrow arteries.
تدمد: 1078-5884
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::610451c13e84922afa1c668b021acaac
https://pubmed.ncbi.nlm.nih.gov/7633984
حقوق: OPEN
رقم الأكسشن: edsair.doi.dedup.....610451c13e84922afa1c668b021acaac