Validation of PTV margin for Gamma Knife Icon frameless treatment using a PseudoPatient® Prime anthropomorphic phantom

التفاصيل البيبلوغرافية
العنوان: Validation of PTV margin for Gamma Knife Icon frameless treatment using a PseudoPatient® Prime anthropomorphic phantom
المؤلفون: Evangelos Pappas, E. Zoros, G. Kalaitzakis, K. Zourari, T. Boursianis, Yao Ding, Dershan Luo, Eun Young Han, Zhifei Wen, Tina Marie Briere, Jihong Wang, Parmeswaran Diagaradjane
المصدر: Journal of Applied Clinical Medical Physics
بيانات النشر: Wiley, 2020.
سنة النشر: 2020
مصطلحات موضوعية: gel, anthropomorphic phantom, Planning target volume, Dose profile, Gamma knife, Radiosurgery, PTV, Imaging phantom, 030218 nuclear medicine & medical imaging, Motion, 03 medical and health sciences, 0302 clinical medicine, Planned Dose, Technical Note, Gamma Knife Icon, Humans, Radiology, Nuclear Medicine and imaging, Radiometry, Instrumentation, Mathematics, Radiation, Brain Neoplasms, Phantoms, Imaging, business.industry, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Ptv margin, 030220 oncology & carcinogenesis, Ionization chamber, Anthropomorphic phantom, Technical Notes, Nuclear medicine, business
الوصف: The Gamma Knife Icon allows the treatment of brain tumors mask‐based single‐fraction or fractionated treatment schemes. In clinic, uniform axial expansion of 1 mm around the gross tumor volume (GTV) and a 1.5 mm expansion in the superior and inferior directions are used to generate the planning target volume (PTV). The purpose of the study was to validate this margin scheme with two clinical scenarios: (a) the patient’s head remaining right below the high‐definition motion management (HDMM) threshold, and (b) frequent treatment interruptions followed by plan adaptation induced by large pitch head motion. A remote‐controlled head assembly was used to control the motion of a PseudoPatient® Prime head phantom; for dosimetric evaluations, an ionization chamber, EBT3 films, and polymer gels were used. These measurements were compared with those from the Gamma Knife plan. For the absolute dose measurements using an ionization chamber, the percentage differences for both targets were less than 3.0% for all scenarios, which was within the expected tolerance. For the film measurements, the two‐dimensional (2D) gamma index with a 2%/2 mm criterion showed the passing rates of ≥87% in all scenarios except the scenario 1. The results of Gel measurements showed that GTV (D100) was covered by the prescription dose and PTV (D95) was well above the planned dose by up to 5.6% and the largest geometric PTV offset was 0.8 mm for all scenarios. In conclusion, the current margin scheme with HDMM setting is adequate for a typical patient’s intrafractional motion.
تدمد: 1526-9914
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::9149a5b05771c2b83369057c59623c66
https://doi.org/10.1002/acm2.12997
حقوق: OPEN
رقم الأكسشن: edsair.doi.dedup.....9149a5b05771c2b83369057c59623c66
قاعدة البيانات: OpenAIRE