Review on cochlear implant electrode array tip fold-over and scalar deviation

التفاصيل البيبلوغرافية
العنوان: Review on cochlear implant electrode array tip fold-over and scalar deviation
المؤلفون: Claude Jolly, Anandhan Dhanasingh
المصدر: Journal of Otology
Journal of Otology, Vol 14, Iss 3, Pp 94-100 (2019)
بيانات النشر: Chinese PLA General Hospital, 2019.
سنة النشر: 2019
مصطلحات موضوعية: Materials science, medicine.medical_treatment, Scalar (mathematics), Straight electrode, Review Article, behavioral disciplines and activities, Electrode insertion, 03 medical and health sciences, 0302 clinical medicine, Cochlear implant, Electrode array, medicine, otorhinolaryngologic diseases, In patient, 030223 otorhinolaryngology, Pre-curved electrode, lcsh:Otorhinolaryngology, lcsh:RF1-547, Clinical Practice, Otorhinolaryngology, Electrode, Electrode tip fold-over, sense organs, Scalar deviation, 030217 neurology & neurosurgery, psychological phenomena and processes, Biomedical engineering
الوصف: Objective: Determine the occurrence rate of cochlear implant (CI) electrode tip fold-over and electrode scalar deviation as reported in patient cases with different commercial electrode types. Data-sources: PubMed search for identifying peer-reviewed articles published till 2018 on CI electrode tip fold-over and scalar deviation. Key-words for searching were “Cochlear electrode tip fold-over”, “Cochlear electrode scalar position” and “Cochlear electrode scalar location”. Articles-selection: Only if electrode related issues were investigated in patient cases. 38 articles met the inclusion-criteria. Results: 13 articles on electrode tip fold-over issue covering 3177 implanted ears, out of which 50 ears were identified with electrode tip fold-over with an occurrence rate of 1.57%. Out of 50 ears, 43 were implanted with pre-curved electrodes and the remaining 7 with lateral-wall electrodes. One article reported on both tip fold-over and scalar deviation. 26 articles reported on the electrode scalar deviation covering an overall number of 2046 ears out of which, 458 were identified with electrode scalar deviation at a rate of 22.38%. After removing the studies that did not report on the number of electrodes per electrode type, it was 1324 ears implanted with pre-curved electrode and 507 ears with lateral-wall electrode. Out of 1324 pre-curved electrode implanted ears, 424 were reported with scalar deviation making an occurrence rate of 32%. Out of 507 lateral-wall electrode implanted ears, 43 were associated with scalar deviation at an occurrence rate of 6.7%. Conclusion: This literature review revealing the fact of higher rate of electrode insertion trauma associated with pre-curved electrode type irrespective of CI brand is one step closer to obsolete it from the clinical practice in the interest of patient's cochlear health. Keywords: Electrode tip fold-over, Scalar deviation, Pre-curved electrode, Straight electrode
اللغة: English
تدمد: 2524-1753
1672-2930
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::a1858f0e4e333018ffe430e14adf0ca4
http://europepmc.org/articles/PMC6712287
حقوق: OPEN