Audit of near patient cholesterol testing in occupational health clinics

التفاصيل البيبلوغرافية
العنوان: Audit of near patient cholesterol testing in occupational health clinics
المؤلفون: H. J. Turner, R. D. O. Cooper, J. M. Stuart, P. A. Laite, F. A. Majeed
المصدر: Occupational Medicine. 43:23-26
بيانات النشر: Oxford University Press (OUP), 1993.
سنة النشر: 1993
مصطلحات موضوعية: medicine.medical_specialty, business.industry, Hypercholesterolemia, Occupational Health Services, Public Health, Environmental and Occupational Health, Reproducibility of Results, Audit, Laboratory results, Ambulatory Care Facilities, Sensitivity and Specificity, Occupational safety and health, Confidence interval, Occupational medicine, Cholesterol, Evaluation Studies as Topic, Emergency medicine, External quality assessment, medicine, Blood cholesterol, Humans, Colorimetry, Cholesterol testing, business, Blood Chemical Analysis
الوصف: An audit of near patient cholesterol testing was carried out in occupational health clinics. The aims were to examine the statistical agreement between Reflotron and laboratory measurements of blood cholesterol and to formulate a policy for the use of Reflotrons in cholesterol testing. Three hundred and fifty-two staff members attending occupational health clinics over a period of 10 months had blood taken for both Reflotron and laboratory measurements. The correlation between the two methods was 0.95. The Reflotron had a negative bias compared to the laboratory, with the mean difference between the two methods of measurement being -0.21 mmol/l (95 per cent confidence interval -0.18 to -0.25 mmol/l). Despite the high correlation coefficient and small mean difference, the scatter of Reflotron-laboratory differences was broad, with 95 per cent of the differences lying in the range of 0.95 mmol/l below to 0.52 mmol/l above the laboratory result. For Reflotron results of 5.50 mmol/l and greater, the sensitivity and specificity of the Reflotron in detecting subjects with laboratory cholesterol levels greater than 6.5 mmol/l were 100 per cent and 70 per cent respectively. The laboratory participated in two external quality assessment schemes for cholesterol testing during the course of the audit and all the results of these fell within the acceptable limits. The audit demonstrated that the Reflotron was too imprecise to be used to give accurate measurements of blood cholesterol. However, providing a suitable Reflotron result above which patients were sent for confirmatory laboratory testing was selected, it was an acceptable screening device in the detection of hypercholesterolaemia. Other Reflotron users should consider carrying out similar audits.
تدمد: 1471-8405
0962-7480
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::b81adb0385da84c04d87f30ba517adcd
https://doi.org/10.1093/occmed/43.1.23
رقم الأكسشن: edsair.doi.dedup.....b81adb0385da84c04d87f30ba517adcd
قاعدة البيانات: OpenAIRE