Sound analysis of the magnetically levitated left ventricular assist device HeartMate 3™

التفاصيل البيبلوغرافية
العنوان: Sound analysis of the magnetically levitated left ventricular assist device HeartMate 3™
المؤلفون: Daniel Karlsson, Fredrik Gustafsson, Hans Granfeldt, Henrik Casimir Ahn, Göran Dellgren, Michael Roth, Laila Hubbert, Per Sundbom
بيانات النشر: Linköpings universitet, Avdelningen för kardiovaskulär medicin, 2019.
سنة النشر: 2019
مصطلحات موضوعية: Male, Computer science, medicine.medical_treatment, Acoustics, Heart Ventricles, Biomedical Engineering, Magnetometry, Medicine (miscellaneous), Bioengineering, Spectral analysis, 030204 cardiovascular system & hematology, HeartMate 3, Biomaterials, sound, 03 medical and health sciences, 0302 clinical medicine, Biomedicinsk laboratorievetenskap/teknologi, medicine, left ventricular assist device, Humans, Biomedical Laboratory Science/Technology, Sound (geography), Heart Failure, geography, geography.geographical_feature_category, Spectrum Analysis, General Medicine, Equipment Design, Middle Aged, Quality Improvement, Equipment Failure Analysis, Sound, 030228 respiratory system, Ventricular assist device, Sound analysis, Female, Heart-Assist Devices
الوصف: Introduction: The HeartMate 3™ has shown lower rates of adverse events compared to previous devices due to the design and absence of mechanical bearings. For previous devices, sound analysis emerged as a way to assess pump function. The aims of this study were to determine if sound analysis can be applied to the HeartMate 3 in vivo and in vitro and to evaluate an electronic stethoscope. Method: Sound recordings were performed with microphones and clinical accessible electronic stethoscope. The recordings were studied in both the time and the frequency domains. Recordings from four patients were performed to determine if in vivo and in vitro recordings are comparable. Results: The results show that it is possible to detect sound from HeartMate 3 and the sound spectrum is clear. Pump frequency and frequency of the pulsatile mode are easily determined. Frequency spectra from in vitro and in vivo recordings have the same pattern, and the major proportion (96.7%) of signal power is located at the pump speed frequency ±40 Hz. The recordings from the patients show low inter-individual differences except from location of peaks originating from pump speed and harmonics. Electronic stethoscopes could be used for sound recordings, but the dedicated equipment showed a clearer sound spectrum. Discussion: The results show that acoustic analysis can also be performed with the HeartMate 3 and that in vivo and in vitro sound spectrum is similar. The frequency spectra are different from previous devices, and methods for assessing pump function or thrombosis need further evaluation.
وصف الملف: application/pdf
اللغة: English
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::cc798de25717f976ce3421f50354a3c7
http://urn.kb.se/resolve?urn=urn:nbn:se:liu:diva-161898
حقوق: OPEN
رقم الأكسشن: edsair.doi.dedup.....cc798de25717f976ce3421f50354a3c7
قاعدة البيانات: OpenAIRE