Comparison of the anchoring strengths of the Kopans and Hawkins II needle-hook-wire systems

التفاصيل البيبلوغرافية
العنوان: Comparison of the anchoring strengths of the Kopans and Hawkins II needle-hook-wire systems
المؤلفون: G F Splittgerber, W S Goell, D J Czarnecki, D L Berridge
المصدر: Radiology. 183(2)
سنة النشر: 1992
مصطلحات موضوعية: Adult, Aged, 80 and over, Hook, business.industry, Tension (physics), Anchoring, Hook wire, Anatomy, Middle Aged, Transverse plane, Needles, Coronal plane, Medicine, Humans, Radiology, Nuclear Medicine and imaging, Female, Composite material, business, Mastectomy, Aged, Mammography
الوصف: To determine whether needle-hook-wire anchoring strength is a factor in accidental repositioning of a hook wire, 5-cm Kopans and Hawkins II hook wires were placed axially (n = 9) and coronally (n = 5) in mastectomy specimens. The tension required to pull the hook wires 1, 2, 3, and 5 cm and to increase the barb angle to 90 degrees and 180 degrees were compared. The average anchoring strength of the Hawkins II hook wire was greater than that of the Kopans hook wire, by 2.6 times (P less than .001; all distances, both angles) with the wires in the axial plane and by 1.6 times (P less than .06; intervals and angles combined) in the coronal plane.
تدمد: 0033-8419
URL الوصول: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::d9a81e97c523ad405eee68f96bcd83cf
https://pubmed.ncbi.nlm.nih.gov/1523330
رقم الأكسشن: edsair.doi.dedup.....d9a81e97c523ad405eee68f96bcd83cf
قاعدة البيانات: OpenAIRE