Feasibility of using a non-sedation protocol for evaluation of neonatal congenital heart disease by using a 16-cm wide-detector computed tomography with a low radiation dose: preliminary experience from a single pediatric medical center

التفاصيل البيبلوغرافية
العنوان: Feasibility of using a non-sedation protocol for evaluation of neonatal congenital heart disease by using a 16-cm wide-detector computed tomography with a low radiation dose: preliminary experience from a single pediatric medical center
المؤلفون: Chen, Guo, Yu-Jie, Liu, Ai-Min, Sun, Rong-Zhen, Ouyang, Li-Wei, Hu, Wei-Hui, Xie, Hai-Sheng, Qiu, Hai-Bo, Zhang, Zhuo-Ming, Xu, Jia-Qi, Sun, Yu-Min, Zhong, Qian, Wang
المصدر: The international journal of cardiovascular imaging. 37(7)
سنة النشر: 2020
مصطلحات موضوعية: Heart Defects, Congenital, Predictive Value of Tests, Infant, Newborn, Feasibility Studies, Humans, Prospective Studies, Child, Coronary Angiography, Radiation Dosage, Tomography, X-Ray Computed
الوصف: The aim of this study is to explore the feasibility of using a non-sedation protocol for the evaluation of neonatal congenital heart disease by using 16-cm wide-detector CT with a low radiation dose. Thirty-four neonates (group 1) were enrolled to undergo cardiac CT without sedation between August 2018 and March 2019. The control group (group 2) comprising 20 inpatient neonates was sedated. Cardiac CT was performed using 16-cm area detector 320-row CT with free breathing and prospective ECG-triggering scan mode. The examination completion time, radiation dose, and image quality were compared between the groups. The results of cardiac CT for patients in group 1 who underwent surgery were compared with surgical findings. Intergroup differences in body weight, age, examination completion time, radiation dose, and image quality evaluation were not significant. There was no significant difference in oxygen saturation before and after the examination in group 1. In all, 98 separate cardiovascular abnormalities in 27 group 1 patients were confirmed using surgical reports. The overall sensitivity, specificity, positive predictive value, and negative predictive value of cardiac CT were 94.90%, 100.0%, 100.0%, and 98.53%. The non-sedation protocol can be applied in neonates with congenital heart disease by using 16-cm wide-detector CT with a low radiation dose. Based on the image quality obtained, non-sedative examination did not extend the examination completion time and helped avoid the possible side effects of sedative drugs.