Study on temperature coefficient of CdTe detector used for X-rays detection

التفاصيل البيبلوغرافية
العنوان: Study on temperature coefficient of CdTe detector used for X-rays detection
المؤلفون: Guo, Si-Ming, Wu, Jin-Jie, Zhang, Jian, Li, Xu-Fang, Liu, Cong-Zhan, Zhang, Shuai, Li, Cheng-Ze, Huo, Bin-Bin, Liao, Zhen-Yu
سنة النشر: 2016
المجموعة: Physics (Other)
مصطلحات موضوعية: Physics - Instrumentation and Detectors
الوصف: The temperature of the working environment is a key factor in determining the properties of semiconductor detectors, and it affects the absolute accuracy and stability of the standard detector. In order to determine the temperature coefficient of CdTe detector used for X-rays detection, a precise temperature control system was designed. In this experiment, detectors and radiographic source were set inside the thermostat with temperature of 0-40 Celsius degree, so that the temperature can be regulated for the test of the temperature coefficient of CdTe detector. Studies had shown that, with the increase of the temperature, the energy resolution and detection efficiency of the CdTe detector would deteriorate, and under 10 Celsius degree the detectors have better performance with the 8 keV X-rays.
Comment: 8 pages, 7 figures
نوع الوثيقة: Working Paper
URL الوصول: http://arxiv.org/abs/1611.06747
رقم الأكسشن: edsarx.1611.06747
قاعدة البيانات: arXiv