SiPM proton irradiation for application in cosmic space

التفاصيل البيبلوغرافية
العنوان: SiPM proton irradiation for application in cosmic space
المؤلفون: Mianowski, S., Borowicz, D. M., Brylew, K., Dziedzic, A., Grodzicka-Kobylka, M., Korgul, A., Krakowiak, M., Mianowska, Z., Molokanov, A. G., Moszynski, M., Mytsin, G. V., Rybka, D., Shipulin, K., Szczesniak, T.
المصدر: 2020 JINST 15 P03002
سنة النشر: 2020
المجموعة: High Energy Physics - Experiment
Physics (Other)
مصطلحات موضوعية: Physics - Instrumentation and Detectors, High Energy Physics - Experiment
الوصف: This paper presents the results of the proton irradiation of silicon photomulipliers (SiPMs) by mono-energetic 170 MeV protons with fluence up to 4.6$\times$10$^{9}$ particles/cm$^2$. In our work, three types of silicon photodetectors from Hamamatsu with areas 3$\times$3 mm$^2$ and different subpixel sizes of 25$\times$25 $\mu$m$^2$, 50$\times$50 $\mu$m$^2$, and 75$\times$75 $\mu$m$^2$ were used. The changes in the SiPM dark count rate (DCR) spectrum before and after irradiation in temperatures in the range of 20 $^\circ$C to -65 $^\circ$C are presented. The influence of the DCR changes on the energy resolution of the 662 keV gamma line from the $^{137}$Cs for a non-irradiated GAGG:Ce (1$\%$) scintillator is investigated. The time period of usability of the SiPM detector irradiated by protons in cosmic space was estimated.
نوع الوثيقة: Working Paper
DOI: 10.1088/1748-0221/15/03/P03002
URL الوصول: http://arxiv.org/abs/2003.09731
رقم الأكسشن: edsarx.2003.09731
قاعدة البيانات: arXiv
الوصف
DOI:10.1088/1748-0221/15/03/P03002