Light Detection System and Time Resolution of the NA62 RICH

التفاصيل البيبلوغرافية
العنوان: Light Detection System and Time Resolution of the NA62 RICH
المؤلفون: Anzivino, G., Barbanera, M., Bizzeti, A., Brizioli, F., Bucci, F., Cassese, A., Cenci, P., Ciaranfi, R., Duk, V., Engelfried, J., Estrada-Tristan, N., Iacopini, E., Imbergamo, E., Latino, G., Lenti, M., Lollini, R., Lubrano, P., Piandani, R., Pepe, M., Piccini, M., Sergi, A., Turisini, M., Volpe, R.
سنة النشر: 2020
المجموعة: High Energy Physics - Experiment
Physics (Other)
مصطلحات موضوعية: Physics - Instrumentation and Detectors, High Energy Physics - Experiment
الوصف: A large RICH detector is used in NA62 to suppress the muon contamination in the charged pion sample by a factor of 100 in the momentum range between 15 and 35 GeV/c. Cherenkov light is collected by 1952 photomultipliers placed at the upstream end. In this paper the characterization of the photomultipliers and the dedicated Frontend and Data Acquisition electronics are described, the time resolution and the light detection efficiency measurement are presented.
Comment: 15 pages, 11 figures
نوع الوثيقة: Working Paper
DOI: 10.1088/1748-0221/15/10/P10025
URL الوصول: http://arxiv.org/abs/2009.07581
رقم الأكسشن: edsarx.2009.07581
قاعدة البيانات: arXiv
الوصف
DOI:10.1088/1748-0221/15/10/P10025