Beat-to-Beat Fetal Heart Rate Analysis Using Portable Medical Device and Wavelet Transformation Technique

التفاصيل البيبلوغرافية
العنوان: Beat-to-Beat Fetal Heart Rate Analysis Using Portable Medical Device and Wavelet Transformation Technique
المؤلفون: Farahi, Maria, Casals, Alicia, Sarrafzadeh, Omid, Zamani, Yasaman, Ahmadi, Hooran, Behbood, Naeimeh, Habibian, Hessam
المصدر: Heliyon 8(12), E12655, 2022
سنة النشر: 2021
المجموعة: Quantitative Biology
مصطلحات موضوعية: Quantitative Biology - Quantitative Methods, 92C55
الوصف: A beat-to-beat Tele-fetal Monitoring and comparison with clinical data are studied with a wavelet transformation approach. Tele-fetal monitoring is a big progress toward a wearable medical device for a pregnant woman capable of obtaining prenatal care at home. We apply a wavelet transformation algorithm for fetal cardiac monitoring using a portable fetal Doppler medical device. Choosing an appropriate mother wavelet, 85 different mother wavelets are investigated. The efficiency of the proposed method is evaluated using two data sets including public and clinical. From publicly available data on PhysioBank, and simultaneous clinical measurement, we prove that the comparison between obtained fetal heart rate by the algorithm and the baselines yields a promising accuracy beyond 95%. Finally, we conclude that the proposed algorithm would be a robust technique for any similar tele-fetal monitoring approach.
Comment: 12 pages, 12 figures
نوع الوثيقة: Working Paper
DOI: 10.1016/j.heliyon.2022.e12655
URL الوصول: http://arxiv.org/abs/2103.01014
رقم الأكسشن: edsarx.2103.01014
قاعدة البيانات: arXiv
الوصف
DOI:10.1016/j.heliyon.2022.e12655