Ortho-positronium Lifetime For Soft-tissue Classification

التفاصيل البيبلوغرافية
العنوان: Ortho-positronium Lifetime For Soft-tissue Classification
المؤلفون: Avachat, Ashish V., Mahmoud, Kholod H., Leja, Anthony G., Xu, Jiajie J., Anastasio, Mark A., Sivaguru, Mayandi, Di Fulvio, Angela
المصدر: Scientific Reports 14.1 (2024): 21155
سنة النشر: 2024
المجموعة: Physics (Other)
مصطلحات موضوعية: Physics - Instrumentation and Detectors
الوصف: The objective of this work is to showcase the ortho-positronium lifetime as a probe for soft-tissue characterization. We employed positron annihilation lifetime spectroscopy to experimentally measure the three components of the positron annihilation lifetime para-positronium (p-Ps), positron, and ortho-positronium (o-Ps) for three types of porcine, non-fixated soft tissues ex vivo: adipose, hepatic, and muscle.Then, we benchmarked our measurements with X-ray phase-contrast imaging, which is the current state-of-the-art for soft-tissue analysis. We found that the o-Ps lifetime in adipose tissues (2.54+/-0.12) ns was approximately 20\% longer than in hepatic (2.04+/-0.09 ns) and muscle (2.03+/-0.12 ns) tissues .In addition, the separation between the measurements for adipose tissue and the other tissues was better from o-Ps lifetime measurement than from X-ray phase-contrast imaging. This experimental study proved that the o-Ps lifetime is a viable non-invasive probe for characterizing and classifying the different soft tissues. Specifically, o-Ps lifetime as a soft-tissue characterization probe had a strong sensitivity to the lipid content that can be potentially implemented in commercial positron emission tomography scanners that feature list-mode data acquisition.
نوع الوثيقة: Working Paper
DOI: 10.1038/s41598-024-71695-7
URL الوصول: http://arxiv.org/abs/2409.13102
رقم الأكسشن: edsarx.2409.13102
قاعدة البيانات: arXiv
الوصف
DOI:10.1038/s41598-024-71695-7