الوصف: Peripheral nerve stimulation (PNS) is a useful treatment for chronic pain, but it can cause damage depending on its application site. Here, we describe the case of a 54-year-old man who underwent PNS for brachial plexopathy in 2015. One lead was implanted on the left medial cord to stimulate the medial antebrachial cutaneous nerve, and the other was implanted on the radial nerve to stimulate the posterior antebrachial cutaneous nerve. Both leads were inserted near the shoulder joint but did not cross it. Before PNS, the patient did not move his shoulder and elbow because of severe pain, but the treatment greatly alleviated this pain. Twenty months after the operation, both leads were fractured, and the severe pain returned. Repetitive motion near the joint was closely related to the lead fractures. In conclusion, large joints as the insertion sites of PNS leads should be avoided to prevent lead fractures.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2005-6419
2005-7563
46616403
Relation: http://ekja.org/upload/pdf/kja-d-17-00023.pdf; https://doaj.org/toc/2005-6419; https://doaj.org/toc/2005-7563
DOI: 10.4097/kja.d.17.00023
URL الوصول: https://doaj.org/article/0d46616403bf4dc289cff08ec5e8ba5c
رقم الأكسشن: edsdoj.0d46616403bf4dc289cff08ec5e8ba5c
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:20056419
20057563
46616403
DOI:10.4097/kja.d.17.00023