دورية أكاديمية

Comparison of Labsan Tricell-1000 and Dirui FUS-2000 automated urine analyzers with manual microscopy

التفاصيل البيبلوغرافية
العنوان: Comparison of Labsan Tricell-1000 and Dirui FUS-2000 automated urine analyzers with manual microscopy
المؤلفون: Sedat Abusoglu, Halil Guven, Busra Ecer, Ahmet Emre Yorulmaz, Abdullah Sivrikaya, Fatma Humeyra Yerlikaya Aydemir, Ali Unlu, Gulsum Abusoglu, Muhittin Abdulkadir Serdar
المصدر: Practical Laboratory Medicine, Vol 39, Iss , Pp e00386- (2024)
بيانات النشر: Elsevier, 2024.
سنة النشر: 2024
المجموعة: LCC:Medicine (General)
LCC:Chemistry
مصطلحات موضوعية: Urinalysis, Concordance, Kappa, Particle recognition, Medicine (General), R5-920, Chemistry, QD1-999
الوصف: Objectives: Urinalysis is a first-line test for screening for urinary tract infection. Several devices performing strip and sediment analysis have been introduced. The aim of this study was to compare the performance of Labsan Tricell-1000 and Dirui FUS-2000 automated urine analyzers with manual microscopy. Methods: 463 urine samples were analyzed. Digital image processing and particle recognition automatically display the cells in a flowing sheath fluid mixed monolayer urine sample, take the pictures of particles via digital camera, analyse these pics with a particle recognition software, transfer images of the formed elements to the screen and allow well-trained personnel to select, reclassify or remove them. Manual microscopy was used for comparison. Results: Agreement between Tricell-100 and manual microscopy was very good for RBC (ϰ = 0.80), and WBC (ϰ = 0.83); good for CaOx (ϰ = 0.69), SEC (ϰ = 0.80), YLC (ϰ = 0.72), HC (0.69) and LC (ϰ = 0.64); moderate for BAC (ϰ = 0.51), APC (ϰ = 0.43) and MT (ϰ = 0.55); fair for GC (ϰ = 0.39) and RTEC (ϰ = 0.32). Conclusions: Labsan Trion TriCell-1000 demonstrated satisfactory performance and can be used in routine urinalysis. In the case of low counts of RBC, presence of yeast, crystal, casts or cell clumping in urine sediment, characterization of urine particles should be performed by manual microscopy.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2352-5517
Relation: http://www.sciencedirect.com/science/article/pii/S2352551724000325; https://doaj.org/toc/2352-5517
DOI: 10.1016/j.plabm.2024.e00386
URL الوصول: https://doaj.org/article/0dfd7beef9d94182816f412492c36439
رقم الأكسشن: edsdoj.0dfd7beef9d94182816f412492c36439
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:23525517
DOI:10.1016/j.plabm.2024.e00386