Biomechanical Comparison of Medio-Plantar and Plantar Plate Fixation for First Tarsometatarsal Joint Arthrodesis

التفاصيل البيبلوغرافية
العنوان: Biomechanical Comparison of Medio-Plantar and Plantar Plate Fixation for First Tarsometatarsal Joint Arthrodesis
المؤلفون: Kajetan Klos, Paul Simons, Pauline Schopp, Philipp Schenk, Felix C. Kohler, Akram Uddin, Edgar K. Roth, Uta Biedermann, Gunther O. Hofmann, Mark Lenz
المصدر: Journal of Clinical Medicine, Vol 12, Iss 12, p 3896 (2023)
بيانات النشر: MDPI AG, 2023.
سنة النشر: 2023
المجموعة: LCC:Medicine
مصطلحات موضوعية: Lapidus bunionectomy, bunion, hallux abducto valgus, locking plate, Medicine
الوصف: Plantar plate positioning has been demonstrated as biomechanically superior. However, some operators remain resentful about the morbidity of the surgical approach. To provide improved plate fixation for first tarsometatarsal joint arthrodesis with respect to the tibialis anterior tendon, a medio-plantar plate was developed. The purpose of this biomechanical study was to compare its construct stability to that of a plantar plate construct. Twelve pairs of fresh frozen human specimens were used in a matched pair test. Each pair was fixed with a 4 mm compression screw and either a plantar locking plate or a medio-plantar locking plate. A cantilever beam test was performed in dorsiflexion. Before and after cyclic loading (5000 cycles; 40 N), bending stiffness and relative movements at the joint space were monitored in a quasi-static test including optical motion tracking. Maximum load and bending moment to failure were investigated in a load-to-failure ramp test. The bending stiffness of both groups did not significantly differ before (plantar 49.9 N/mm ± 19.2; medio-plantar 53.9 N/mm ± 25.4, p = 0.43) or after (plantar 24.4 N/mm ± 9.7; medio-plantar 35.3 N/mm ± 22.0, p = 0.08) cyclic loading but decreased significantly in both groups (p < 0.01) after cyclic loading. Relative movement increased significantly during cyclic testing in both groups (p < 0.01) but did not differ significantly between the groups before (p = 0.29) or after (p = 0.16) cyclic loading. Neither load nor bending moment to failure were significantly different (plantar 225 N ± 78, 10.8 Nm; medio-plantar 210 N ± 86, 10.1 Nm, p = 0.61). Both plate constructs provided equivalent construct stability, both being well suited for Lapidus arthrodesis.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2077-0383
Relation: https://www.mdpi.com/2077-0383/12/12/3896; https://doaj.org/toc/2077-0383
DOI: 10.3390/jcm12123896
URL الوصول: https://doaj.org/article/c0e25d959e634abaae1bb70bf83d3f80
رقم الأكسشن: edsdoj.0e25d959e634abaae1bb70bf83d3f80
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:20770383
DOI:10.3390/jcm12123896