Accuracy Evaluation of Carotid-Femoral Pulse Wave Velocity Estimated by Smart Terminal Watch

التفاصيل البيبلوغرافية
العنوان: Accuracy Evaluation of Carotid-Femoral Pulse Wave Velocity Estimated by Smart Terminal Watch
المؤلفون: Ningling Sun, Luyan Wang, Yang Xi, Hongyi Wang, Fan Yang, Yuanyuan Chen, Jing Liu, Yuxian Cui, Zhechun Zeng
المصدر: Frontiers in Cardiovascular Medicine, Vol 9 (2022)
بيانات النشر: Frontiers Media S.A., 2022.
سنة النشر: 2022
المجموعة: LCC:Diseases of the circulatory (Cardiovascular) system
مصطلحات موضوعية: cervical-femoral pulse wave velocity, arterial stiffness, smartwatch, accuracy, Complior arteriosclerosis analyzer, Diseases of the circulatory (Cardiovascular) system, RC666-701
الوصف: To evaluate the accuracy of the smartwatch in estimating carotid-femoral pulse wave velocity (cfPWV). A cohort of gender-matched volunteers aged 18–80 years were recruited. At the sitting and supine positions, cfPWV was measured alternately by smartwatch and CompliorAnalyse, for each participant, and nine sets of data were collected from each participant with a 60 s interval between measurements. The accuracy of cfPWV measurement for smartwatches was assessed using mean error (ME) and mean absolute error (MAE), while the consistency of the two methods was assessed using the Bland-Altman analysis and concordance class correlation. A total of 347 participants were enrolled. The mean cfPWV was 9.01 ± 2.29 m/s measured by CompliorAnalyse and 9.06 ± 1.94 m/s by smartwatch. The consistency correlation coefficient (CCC) was 0.9045 (95% CI 0.8853–0.9206), the ME was 0.046 ± 0.92, and the MAE was 0.66 (95% CI 0.59–0.73). Bland-Altman analysis showed that the error of 95% samples was in the range between −1.77 m/s and 1.86 m/s. The Kappa value of cfPWV greater than 10 m/s was 0.79, the area under the ROC curve was 0.97 (P < 0.001), sensitivity was 0.90, specificity was 0.93, positive predictive value was 0.83 and negative predictive value was 0.96. Smartwatch can accurately estimate cfPWV to evaluate arterial stiffness. This method is simple and feasible and is suitable for people to actively and early monitor vascular elasticity.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2297-055X
Relation: https://www.frontiersin.org/articles/10.3389/fcvm.2022.893557/full; https://doaj.org/toc/2297-055X
DOI: 10.3389/fcvm.2022.893557
URL الوصول: https://doaj.org/article/2152c331f5194f2fbafd1699f1ca6c13
رقم الأكسشن: edsdoj.2152c331f5194f2fbafd1699f1ca6c13
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:2297055X
DOI:10.3389/fcvm.2022.893557