Abstract Background Superiority of noninvasive tripolar concentric ring electrodes over conventional disc electrodes in accuracy of surface Laplacian estimation has been demonstrated in a range of electrophysiological measurement applications. Recently, a general approach to Laplacian estimation for an (n + 1)-polar electrode with n rings using the (4n + 1)-point method has been proposed and used to introduce novel multipolar and variable inter-ring distances electrode configurations. While only linearly increasing and linearly decreasing inter-ring distances have been considered previously, this paper defines and solves the general inter-ring distances optimization problem for the (4n + 1)-point method. Results General inter-ring distances optimization problem is solved for tripolar (n = 2) and quadripolar (n = 3) concentric ring electrode configurations through minimizing the truncation error of Laplacian estimation. For tripolar configuration with middle ring radius αr and outer ring radius r the optimal range of values for α was determined to be 0