Comparison of the New Viscoelastic Coagulation Analyzer ClotPro® With ROTEM® Delta and Conventional Coagulation Tests in Critically Ill Patients With COVID-19

التفاصيل البيبلوغرافية
العنوان: Comparison of the New Viscoelastic Coagulation Analyzer ClotPro® With ROTEM® Delta and Conventional Coagulation Tests in Critically Ill Patients With COVID-19
المؤلفون: Lukas Infanger, Christoph Dibiasi, Eva Schaden, Stefan Ulbing, Marion Wiegele, Conrad Lacom, Johannes Gratz
المصدر: Frontiers in Medicine, Vol 8 (2021)
بيانات النشر: Frontiers Media S.A., 2021.
سنة النشر: 2021
المجموعة: LCC:Medicine (General)
مصطلحات موضوعية: ClotPro®, coagulation, comparison, COVID-19, critically ill, point-of-care, Medicine (General), R5-920
الوصف: Background: Viscoelastic coagulation testing has been suggested to help manage coagulopathy in critically ill patients with COVID-19. However, results from different viscoelastic devices are not readily comparable. ClotPro® is a novel thromboelastometry analyzer offering a wider range of commercially available assays.Methods: We compared the results from ClotPro with results from the well-established ROTEM® Delta device and conventional coagulation tests in critically ill patients with COVID-19.Results: Viscoelastic parameters indicated the presence of a potentially hypercoagulable state in the majority of patients. In up to 95 paired measurements, we found strong correlations between several parameters routinely used in clinical practice: (i) EX test vs. EXTEM CT, A5, A10, MCF, (ii) IN test vs. INTEM A5, A10, MCF, and (iii) FIB test vs. FIBTEM A5, A10, MCF (all R > 0.7 and p < 0.001). In contrast, IN test CT vs. INTEM CT showed only a moderate correlation (R = 0.53 and p < 0.001). Clot strength parameters of both devices exhibited strong correlations with platelet counts and fibrinogen levels (all R > 0.7 and p < 0.001). Divergent correlations of intrinsically activated assays with aPTT and anti-factor Xa activity were visible. Regarding absolute differences of test results, considerable delta occurred in CT, CFT, and clot strength parameters (all p < 0.001) between both devices.Conclusions: Several parameters obtained by ClotPro show strong correlations with ROTEM Delta. Due to weak correlations of intrinsically activated clotting times and considerable absolute differences in a number of parameters, our findings underline the need for device-specific algorithms in this patient cohort.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2296-858X
Relation: https://www.frontiersin.org/articles/10.3389/fmed.2021.777145/full; https://doaj.org/toc/2296-858X
DOI: 10.3389/fmed.2021.777145
URL الوصول: https://doaj.org/article/40ec18bbfa1d479db18509d387bc58b7
رقم الأكسشن: edsdoj.40ec18bbfa1d479db18509d387bc58b7
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:2296858X
DOI:10.3389/fmed.2021.777145