Impact of respiratory motion on lung dose during total marrow irradiation

التفاصيل البيبلوغرافية
العنوان: Impact of respiratory motion on lung dose during total marrow irradiation
المؤلفون: Ayse Gulbin Kavak, Murat Surucu, Kang-Hyun Ahn, Erik Pearson, Bulent Aydogan
المصدر: Frontiers in Oncology, Vol 12 (2022)
بيانات النشر: Frontiers Media S.A., 2022.
سنة النشر: 2022
المجموعة: LCC:Neoplasms. Tumors. Oncology. Including cancer and carcinogens
مصطلحات موضوعية: organ motion, dose delivery, breathing motion, TMI, total marrow irradiation, Neoplasms. Tumors. Oncology. Including cancer and carcinogens, RC254-282
الوصف: We evaluated the impact of respiratory motion on the lung dose during linac-based intensity-modulated total marrow irradiation (IMTMI) using two different approaches: (1) measurement of doses within the lungs of an anthropomorphic phantom using thermoluminescent detectors (TLDs) and (2) treatment delivery measurements using ArcCHECK where gamma passing rates (GPRs) and the mean lung doses were calculated and compared with and without motion. In the first approach, respiratory motions were simulated using a programmable motion platform by using typical published peak-to-peak motion amplitudes of 5, 8, and 12 mm in the craniocaudal (CC) direction, denoted here as M1, M2, and M3, respectively, with 2 mm in both anteroposterior (AP) and lateral (LAT) directions. TLDs were placed in five selected locations in the lungs of a RANDO phantom. Average TLD measurements obtained with motion were normalized to those obtained with static phantom delivery. The mean dose ratios were 1.01 (0.98–1.03), 1.04 (1.01–1.09), and 1.08 (1.04–1.12) for respiratory motions M1, M2, and M3, respectively. To determine the impact of directional respiratory motion, we repeated the experiment with 5-, 8-, and 12-mm motion in the CC direction only. The differences in average TLD doses were less than 1% when compared with the M1, M2, and M3 motions indicating a minimal impact from CC motion on lung dose during IMTMI. In the second experimental approach, we evaluated extreme respiratory motion 15 mm excursion in only the CC direction. We placed an ArcCHECK device on a commercial motion platform and delivered the clinical IMTMI plans of five patients. We compared, with and without motion, the dose volume histograms (DVHs) and mean lung dose calculated with the ArcCHECK-3DVH tool as well as GPR with 3%, 5%, and 10% dose agreements and a 3-mm constant distance to agreement (DTA). GPR differed by 11.1 ± 2.1%, 3.8 ± 1.5%, and 0.1 ± 0.2% with dose agreement criteria of 3%, 5%, and 10%, respectively. This indicates that respiratory motion impacts dose distribution in small and isolated parts of the lungs. More importantly, the impact of respiratory motion on the mean lung dose, a critical indicator for toxicity in IMTMI, was not statistically significant (p > 0.05) based on the Student’s t-test. We conclude that most patients treated with IMTMI will have negligible dose uncertainty due to respiratory motion. This is particularly reassuring as lung toxicity is the main concern for future IMTMI dose escalation studies.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2234-943X
Relation: https://www.frontiersin.org/articles/10.3389/fonc.2022.924961/full; https://doaj.org/toc/2234-943X
DOI: 10.3389/fonc.2022.924961
URL الوصول: https://doaj.org/article/420a7c1ce5e34e899c6be67fc01b8970
رقم الأكسشن: edsdoj.420a7c1ce5e34e899c6be67fc01b8970
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:2234943X
DOI:10.3389/fonc.2022.924961