Establishment of high-throughput liquid chromatography tandem mass spectrometry method for determination of 53 per- and polyfluoroalkyl substances in serum

التفاصيل البيبلوغرافية
العنوان: Establishment of high-throughput liquid chromatography tandem mass spectrometry method for determination of 53 per- and polyfluoroalkyl substances in serum
المؤلفون: Zheng WANG, Boya ZHANG, Jiming ZHANG, Chao FENG, Yuanjie LIN, Chunhua WU, Dasheng LU, Zhijun ZHOU
المصدر: 环境与职业医学, Vol 41, Iss 4, Pp 375-383 (2024)
بيانات النشر: Editorial Committee of Journal of Environmental and Occupational Medicine, 2024.
سنة النشر: 2024
المجموعة: LCC:Medicine (General)
LCC:Toxicology. Poisons
مصطلحات موضوعية: per- and polyfluoroalkyl substances, serum, high resolution mass spectrometry, solid phase extraction, high throughput, Medicine (General), R5-920, Toxicology. Poisons, RA1190-1270
الوصف: BackgroundPer- and polyfluoroalkyl substances (PFAS) are a class of persistent organic pollutants that possess potential toxicity to the human body. The production and utilization of diverse emerging PFAS have resulted in widespread human exposure. Therefore, it is imperative to establish a quantitative methodology encompassing a wide range of PFAS for a comprehensive assessment of human exposure to these compounds. ObjectiveTo establish a high-throughput quantitative method for the simultaneous determination of 53 PFAS in human serum based on ultra-high-performance liquid chromatography-Q Exactive high resolution mass spectrometry (UPLC-Q Exactive HRMS). MethodsThe extraction recoveries of hydrophilic-lipophilic balance (HLB) column, weak anionexchange (WAX) column, and 96-well WAX μElution plate were compared to select the SPE column with the highest recovery. The retention time and peak shape of the target compounds were compared between ACQUITY UPLC BEH C18 column and Accucore aQ column, and the more cost-effective column was chosen. The effects of adding different levels of ammonium formate (0, 2, 5 and 10 mmol·L−1) in mobile phase on peak shape and target response were compared to determine the optimal buffer salt concentration. The optimal spray voltage was obtained by comparing −2 kV and −4 kV. The proposed method was validated from the aspects of selectivity, standard curve, limits of detection, precision, accuracy, and matrix effect. The method was applied to 142 umbilical serum samples. ResultsThe best recovery rate (64%-118%) was achieved by using 96-well WAX μElution plate. The optimal separation and peak shape were obtained by utilizing Accucore aQ column with H2O-methanol (containing 5 mmol·L−1 ammonium formate) as the mobile phase. Less in-source collision and better target response were observed when the spray voltage was set to −2 kV. All target analytes had a good linearity, with R2 > 0.99. The limits of detection ranged from 0.01 to 0.50 μg·L−1, and the recovery ranged from 69% to 127% with the precision less than 26%. A total of 31 PFAS were detected in the 142 actual samples, among which 14 PFAS had a detection frequency over 50%. Perfluorooctanoic acid showed the highest median concentration of 4.16 μg·L−1, followed by 6:2 chlorinated polyfluorinated ether sulfonate and perfluorooctane sulfonates (3.50 μg·L−1 and 1.59 μg·L−1, respectively). ConclusionIn this study, we establish a UPLC-Q Excative HRMS method for simutanious determination of 53 PFAS concentrations in serum. This method has the advantages of wide coverage of PFAS, good selectivity, and easy operation, and is suitable for biological detection with a large sample size.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
Chinese
تدمد: 2095-9982
Relation: http://www.jeom.org/article/cn/10.11836/JEOM23362; https://doaj.org/toc/2095-9982
DOI: 10.11836/JEOM23362
URL الوصول: https://doaj.org/article/52f0861736764ef6a52cf1eda6eb86f7
رقم الأكسشن: edsdoj.52f0861736764ef6a52cf1eda6eb86f7
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:20959982
DOI:10.11836/JEOM23362