الوصف: Study DesignMechanical study.PurposeTo compare the pullout strength of different screw designs and augmentation techniques in an osteoporotic bone model.Overview of LiteratureAdequate bone screw pullout strength is a common problem among osteoporotic patients. Various screw designs and augmentation techniques have been developed to improve the biomechanical characteristics of the bone–screw interface.MethodsPolyurethane blocks were used to mimic human osteoporotic cancellous bone, and six different screw designs were tested. Five standard and expandable screws without augmentation, eight expandable screws with polymethylmethacrylate (PMMA) or calcium phosphate augmentation, and distal cannulated screws with PMMA and calcium phosphate augmentation were tested. Mechanical tests were performed on 10 unused new screws of each group. Screws with or without augmentation were inserted in a block that was held in a fixture frame, and a longitudinal extraction force was applied to the screw head at a loading rate of 5 mm/min. Maximum load was recorded in a load displacement curve.ResultsThe peak pullout force of all tested screws with or without augmentation was significantly greater than that of the standard pedicle screw. The greatest pullout force was observed with 40-mm expandable pedicle screws with four fins and PMMA augmentation. Augmented distal cannulated screws did not have a greater peak pullout force than nonaugmented expandable screws. PMMA augmentation provided a greater peak pullout force than calcium phosphate augmentation.ConclusionsExpandable pedicle screws had greater peak pullout forces than standard pedicle screws and had the advantage of augmentation with either PMMA or calcium phosphate cement. Although calcium phosphate cement is biodegradable, osteoconductive, and nonexothermic, PMMA provided a significantly greater peak pullout force. PMMA-augmented expandable 40-mm four-fin pedicle screws had the greatest peak pullout force.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 1976-1902
1976-7846
Relation: http://www.asianspinejournal.org/upload/pdf/asj-12-3.pdf; https://doaj.org/toc/1976-1902; https://doaj.org/toc/1976-7846
DOI: 10.4184/asj.2018.12.1.3
URL الوصول: https://doaj.org/article/5501080057344bf7b8b0e9f986020686
رقم الأكسشن: edsdoj.5501080057344bf7b8b0e9f986020686
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:19761902
19767846
DOI:10.4184/asj.2018.12.1.3