NORMATIVE VALUES FOR A NEW WATER-PERFUSED HIGH RESOLUTION MANOMETRY SYSTEM

التفاصيل البيبلوغرافية
العنوان: NORMATIVE VALUES FOR A NEW WATER-PERFUSED HIGH RESOLUTION MANOMETRY SYSTEM
المؤلفون: Rogério Mariotto Bitetti da SILVA, Fernando A M HERBELLA, Daniel GUALBERTO
المصدر: Arquivos de Gastroenterologia, Iss 0 (2018)
بيانات النشر: Instituto Brasileiro de Estudos e Pesquisas de Gastroenterologia (IBEPEGE), 2018.
سنة النشر: 2018
المجموعة: LCC:Diseases of the digestive system. Gastroenterology
مصطلحات موضوعية: Transtornos da motilidade esofágica, Manometria, tendências, Tecnologia de baixo custo, Diseases of the digestive system. Gastroenterology, RC799-869
الوصف: ABSTRACT BACKGROUND: Esophageal manometry is the most reliable method to evaluate esophageal motility. High resolution manometry (HRM) provides topographic contour colored plots (Clouse Plots) with simultaneous analysis from the pharynx to the stomach. Both solid state and water-perfused systems are available. OBJECTIVE: This study aims to determinate the normative data for a new water-perfused HRM. METHODS: HRM was made in 32 healthy volunteers after 8 hours fasting. HRM system used consisted of a 24-channel water-perfused catheter (Multiplex, Alacer Biomedica, São Paulo, Brazil). The reusable catheter is made of polyvinyl chloride (PVC) with 4.7 mm of diameter. Side holes connected to pressure transducers are spaced 2 cm for the analysis from the pharynx to the lower esophageal sphincter (LES). Holes are spaced 5 mm and 120° in a spiral disposition in the LES area. The sensors encompass 34 cm in total. Upper esophageal sphincter (UES) parameters studied were basal and relaxation pressures. Esophageal body parameters were distal contractile integral (DCI), distal latency (DL) and break. LES parameters studied were basal pressure, integrated residual pressure (IRP), total and abdominal length. Variables are expressed as mean ± standard deviation, median (interquartile range) and percentiles 5-95th. RESULTS: All volunteers (17 males, aged 22-62 years) completed the study and tolerated the HRM procedure well. Percentiles 5-95th range were calculated: Upper Esophageal Sphincter (UES) basal pressure 16.7-184.37 (mmHg), DL: 6.2-9.1 (s), DCI: 82.72-3836.61 (mmHg.s.cm), break:
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 1678-4219
0004-2803
Relation: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0004-28032018005002103&lng=en&tlng=en; https://doaj.org/toc/1678-4219
DOI: 10.1590/s0004-2803.201800000-40
URL الوصول: https://doaj.org/article/602e98ade9524ffa887853fe40c415e6
رقم الأكسشن: edsdoj.602e98ade9524ffa887853fe40c415e6
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:16784219
00042803
DOI:10.1590/s0004-2803.201800000-40