دورية أكاديمية

Linear-Rotary Position Control System With Enhanced Disturbance Rejection for a Novel Total Artificial Heart

التفاصيل البيبلوغرافية
العنوان: Linear-Rotary Position Control System With Enhanced Disturbance Rejection for a Novel Total Artificial Heart
المؤلفون: Rosario V. Giuffrida, Andreas Horat, Dominik Bortis, Tim Bierewirtz, Krishnaraj Narayanaswamy, Marcus Granegger, Johann W. Kolar
المصدر: IEEE Open Journal of the Industrial Electronics Society, Vol 5, Pp 359-375 (2024)
بيانات النشر: IEEE, 2024.
سنة النشر: 2024
المجموعة: LCC:Electronics
LCC:Industrial engineering. Management engineering
مصطلحات موضوعية: Artificial biological organs, permanent magnet machines, rotating machines, Electronics, TK7800-8360, Industrial engineering. Management engineering, T55.4-60.8
الوصف: A novel implantable total artificial heart, hereinafter referred to as the ShuttlePump, is currently under development in a research collaboration between the Medical University of Vienna, the Power Electronic Systems Laboratory of ETH Zurich and Charite Berlin. Its novel, low-complexity, pulsatile pumping principle requires a specially shaped piston performing a controlled, synchronized linear-rotary motion while providing the necessary hydraulic force and torque. The machine design of the Permanent Magnet Synchronous Machine (PMSM)-based linear-rotary actuator was conducted in previous work of the authors, leading to the construction of a hardware prototype satisfying the application requirements in terms of electromechanical force, torque, power losses, and volume. This article provides the details of the closed-loop linear-rotary position control system required to operate the ShuttlePump. The design of the position control system targets tight reference tracking ($\pm$8 mm linear stroke and continuous rotation) up to an operational frequency of 5 Hz, under the heavy disturbance introduced by the axial hydraulic load force, as high as 45 N. The experimental measurements show successful linear-rotary position tracking under the specified axial load, with a maximum error of 1 mm and 5$^{\circ }$.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2644-1284
Relation: https://ieeexplore.ieee.org/document/10494377/; https://doaj.org/toc/2644-1284
DOI: 10.1109/OJIES.2024.3385865
URL الوصول: https://doaj.org/article/60a27d899f074e0089beff1196ec26a0
رقم الأكسشن: edsdoj.60a27d899f074e0089beff1196ec26a0
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:26441284
DOI:10.1109/OJIES.2024.3385865