دورية أكاديمية

Performance of an extrapolation chamber in computed tomography standard beams

التفاصيل البيبلوغرافية
العنوان: Performance of an extrapolation chamber in computed tomography standard beams
المؤلفون: Maysa Costa Castro, Natállia Fiorini Silva, Linda Viola Ehlin Caldas
المصدر: Brazilian Journal of Radiation Sciences, Vol 8, Iss 3B (2021)
بيانات النشر: Brazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR), 2021.
سنة النشر: 2021
المجموعة: LCC:Science
مصطلحات موضوعية: Extrapolation chamber, computed tomography, dosimetry., Science
الوصف: Among the medical uses of ionizing radiations, the computed tomography (CT) diagnostic exams are responsible for the highest dose values to the patients. The dosimetry procedure in CT scanner beams makes use of pencil ionization chambers with sensitive volume lengths of 10 cm. The aim of its calibration is to compare the values that are obtained with the instrument to be calibrated and a standard reference system. However, there is no primary standard system for this kind of radiation beam. Therefore, an extrapolation ionization chamber, built at the Calibration Laboratory (LCI), was used to establish a CT primary standard. The objective of this work was to perform some characterization tests (short- and medium-term stabilities, saturation curve, polarity effect and ion collection efficiency) in the standard X-ray beams established for computed tomography at the LCI.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
Portuguese
تدمد: 2319-0612
Relation: https://bjrs.org.br/revista/index.php/REVISTA/article/view/376; https://doaj.org/toc/2319-0612
DOI: 10.15392/bjrs.v8i3B.376
URL الوصول: https://doaj.org/article/700a742c084d4dbdb21213b3d04eee37
رقم الأكسشن: edsdoj.700a742c084d4dbdb21213b3d04eee37
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:23190612
DOI:10.15392/bjrs.v8i3B.376