Background: Atrial AutoCapture™ (ACap™) was a new technological development that confirmed atrial capture by analyzing evoked response (ER) with a new method – paced depolarization integral ER detection – and optimized energy output to changes in the stimulation threshold. The purpose of this study was to evaluate the clinical performance of ACap™ function. Methods: This was a prospective, observational, nonrandomized two-center study. Between November 2008 and August 2014, 102 patients were enrolled from two different institutions. Data were collected by case report forms at enrollment, hospital discharge, and in-office follow-ups scheduled at 1, 2, 3, 6, and 12 months postimplantation. Results: Ambulatory ACap™ function started to become available for 20.6% of patients at 1 day, then progressed to 30.4% at 7 days, 38.6% at 1 month, 41.6% at 2 months, 47.5% at 3 months, 53.5% at 6 months, and 63.4% at 1 year. The cause of the unsuccessful attempts to perform ACap™ threshold was ER/polarization