دورية أكاديمية

Performance of bioelectrical impedance analysis compared to dual X-ray absorptiometry (DXA) in Veterans with COPD

التفاصيل البيبلوغرافية
العنوان: Performance of bioelectrical impedance analysis compared to dual X-ray absorptiometry (DXA) in Veterans with COPD
المؤلفون: Paola N. Cruz Rivera, Rebekah L. Goldstein, Madeline Polak, Antonio A. Lazzari, Marilyn L. Moy, Emily S. Wan
المصدر: Scientific Reports, Vol 12, Iss 1, Pp 1-8 (2022)
بيانات النشر: Nature Portfolio, 2022.
سنة النشر: 2022
المجموعة: LCC:Medicine
LCC:Science
مصطلحات موضوعية: Medicine, Science
الوصف: Abstract We examined the performance of a commercially-available handheld bioimpedance (BIA) device relative to dual X-ray absorptiometry (DXA) to assess body composition differences among Veterans with chronic obstructive pulmonary disease (COPD). Body composition was measured using DXA and BIA (Omron HBF-306C) at a single time point. Correlations between BIA- and DXA-assessed percent fat, fat mass, and fat-free mass were analyzed using Spearman (ρ) and Lin Concordance Correlation Coefficients (ρc). Mean differences in fat mass were visualized using Bland–Altman plots. Subgroup analyses by obesity status (BMI
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2045-2322
Relation: https://doaj.org/toc/2045-2322
DOI: 10.1038/s41598-022-05887-4
URL الوصول: https://doaj.org/article/ff41bd6f0fbc4666a4ed1d7db37a4f76
رقم الأكسشن: edsdoj.ff41bd6f0fbc4666a4ed1d7db37a4f76
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:20452322
DOI:10.1038/s41598-022-05887-4