دورية أكاديمية

Multimodal Simulation of a Novel Device for a Safe and Effective External Ventricular Drain Placement

التفاصيل البيبلوغرافية
العنوان: Multimodal Simulation of a Novel Device for a Safe and Effective External Ventricular Drain Placement
المؤلفون: Giuseppe Emmanuele Umana, Gianluca Scalia, Kaan Yagmurlu, Rosalia Mineo, Simone Di Bella, Matteo Giunta, Angelo Spitaleri, Rosario Maugeri, Francesca Graziano, Marco Fricia, Giovanni Federico Nicoletti, Santino Ottavio Tomasi, Giuseppe Raudino, Bipin Chaurasia, Gianluca Bellocchi, Maurizio Salvati, Domenico Gerardo Iacopino, Salvatore Cicero, Massimiliano Visocchi, Lidia Strigari
المصدر: Frontiers in Neuroscience, Vol 15 (2021)
بيانات النشر: Frontiers Media S.A., 2021.
سنة النشر: 2021
المجموعة: LCC:Neurosciences. Biological psychiatry. Neuropsychiatry
مصطلحات موضوعية: hydrocephalus, shunt, ventricle, cerebral hypertension, ventricular catheter, Neurosciences. Biological psychiatry. Neuropsychiatry, RC321-571
الوصف: BackgroundExternal ventricular drain (EVD) placement is mandatory for several pathologies. The misplacement rate of the EVD varies widely in literature, ranging from 12.3 to 60%. The purpose of this simulation study is to provide preliminary data about the possibility of increasing the safety of one of the most common life-saving procedures in neurosurgery by testing a new device for EVD placement.MethodsWe used a novel guide for positioning the ventricular catheter (patent RM2014A000376). The trajectory was assessed using 25 anonymized head CT scans. The data sets were used to conduct three-dimensional computer-based and combined navigation and augmented reality-based simulations using plaster models. The data set inclusion criteria were volumetric head CT scan, without midline shift, of patients older than 18. Evans’ index was used to quantify the ventricle’s size. We excluded patients with slit ventricles, midline shift, skull fractures, or complex skull malformations. The proximal end of the device was tested on the cadaver.ResultsThe cadaveric tests proved that a surgeon could use the device without any external help. The multimodal simulation showed Kakarla grade 1 in all cases but one (grade 2) on both sides, after right and left EVD placement. The mean Evans’ index was 0.28. The geometric principles that explain the device’s efficacy can be summarized by studying the properties of circumference and chord. The contact occurs, for each section considered, at the extreme points of the chord. Its axis, perpendicular to the plane tangent to the spherical surface at the entry point, corresponds to the direction of entry of the catheter guided by the instrument.ConclusionAccording to our multimodal simulation on cadavers, 3D computer-based simulation, 3D plaster modeling, 3D neuronavigation, and augmented reality, the device promises to offer safer and effective EVD placement. Further validation in future clinical studies is recommended.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 1662-453X
Relation: https://www.frontiersin.org/articles/10.3389/fnins.2021.690705/full; https://doaj.org/toc/1662-453X
DOI: 10.3389/fnins.2021.690705
URL الوصول: https://doaj.org/article/ff8a03e8fb924d06becc75db130d46b5
رقم الأكسشن: edsdoj.ff8a03e8fb924d06becc75db130d46b5
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:1662453X
DOI:10.3389/fnins.2021.690705