مورد إلكتروني
Test and Evaluation Report of the Physio Control Defibrillator/Monitor Model LIFEPAK (Trade Name) 10
العنوان: | Test and Evaluation Report of the Physio Control Defibrillator/Monitor Model LIFEPAK (Trade Name) 10 |
---|---|
المصدر: | DTIC AND NTIS |
بيانات النشر: | 1991-03 |
تفاصيل مُضافة: | ARMY AEROMEDICAL RESEARCH LAB FORT RUCKER AL Haun, Jeffrey D. Licina, Joseph R. Olding, Bill Quattlebaum, Martin |
نوع الوثيقة: | Electronic Resource |
مستخلص: | The Physio Control Defibrillator/Monitor Model LIFEPAK 10 was tested for electromagnetic interference/compatibility in the UH-60A helicopter under the U.S. Army Program for Testing and Evaluation of Equipment for Aeromedical Operations. The tests were conducted using current military and industrial standards and procedures for electromagnetic interference/compatibility and human factors. The LIFEPAK 10 passed the overall evaluation and is validated as compatible with U.S. Army aeromedical aircraft. |
مصطلحات الفهرس: | Medicine and Medical Research, Test Facilities, Equipment and Methods, HUMAN FACTORS ENGINEERING, MILITARY REQUIREMENTS, INDUSTRIES, ELECTROMAGNETIC INTERFERENCE, STANDARDS, ARMY, ARMY AIRCRAFT, AEROMEDICAL EVACUATION., ARMY MEDICINE, LIFEPAK 10, DEFIBRILLATORS, AEROMEDICAL EQUIPMENT, UH-60 AIRCRAFT, PHYSIO CONTROL DEFIBRILLATORS, TEST AND EVALUATION, PE63807A, WU201, AS836, Text |
URL: | |
الإتاحة: | Open access content. Open access content APPROVED FOR PUBLIC RELEASE |
ملاحظة: | text/html English |
أرقام أخرى: | DTICE ADA234593 834260704 |
المصدر المساهم: | From OAIster®, provided by the OCLC Cooperative. |
رقم الأكسشن: | edsoai.ocn834260704 |
قاعدة البيانات: | OAIster |
الوصف غير متاح. |